V1410 / V1412 User Manual
91
Messages – Calibration Mode
Message
Description
FILTERLINE NOT
CONNECTED
The filter line is not connected.
CALIBRATION IN
PROGRESS
The calibration is in progress.
CALIBRATION NOT READY
Not ready to calibrate the EtCO2 Module.
(Run calibration after a while)
CALIBRATION GAS
REMOVEABLE
Remove the EtCO2 calibration gas from the accessory.
CALIBRATION OK
The calibration is complete.
CALIBRATION FAIL
Failed the calibration.
CALIBRATION ERROR -
MEASUREMENT ERROR
Failed to measure the CO
2
gas.
(In case of the wrong setting value)
CALIBRATION ERROR -
WRONG GAS CONNECTION
Not connected to CO
2
gas.
CALIBRATION ERROR -
NO STABLE GAS FLOW
CO
2
GAS pressure is not stable.
CALIBRATION STOP
Calibration has stopped.
(It will not show in normal operation.)