V1410 / V1412 User Manual
100
- DC Offset: -300 mV to 300 mV
- Arrhythmia Display (TAC, BRD, PVC, VTAC, ASY, BGM, TGM, VENT, VFIB, CPT, TPT, MIB,
R on T)
- HR Calculation: 4 – 16 bit
Heart Rate
- Measurement Range: 0 – 300 bpm
- Accuracy: ±2 bpm (measurement value)
- Alarm Range: High Alarm Limit: 0 – 300, OFF
Low Alarm Limit: OFF, 0 – 300
SpO2 (Pulse Oximetery)
- Measurement Range: 0 – 100%
- Accuracy: 70 – 100% SpO2: ± 2%, 50 – 69% SpO2: ± 3%
0 – 49% SpO2: Unspecified
- Alarm Range: High Limit - 0 – 100%, OFF,
Low Limit - OFF, 0 – 100%
- Pulse Range: 0 – 300 bpm
- Pulse Measurement Accuracy: 20 – 240 bpm ±2 bpm
241 – 300 bpm ±3 bpm
NIBP (Non-invasive Blood Pressure Measurement)
- Measurement Method: Oscillometric, Automatic measurement
- Display Value: Systolic, Diastolic, Mean Pressure
- Measurement Range: 0 – 300 mmHg
- Accuracy: ±8 mmHg (systolic, diastolic), ±5 mmHg (mean)
- Alarm Range:
NIBP
[mmHg]
Display
High
Low
Systolic
0 – 275, OFF
OFF, 0 – 275
Diastolic
Mean
- Auto Pressure: Manual, Automatic (1, 2, 3, 4, 5, 10, 15, 20, 30 min. and 1, 2, 4, 8, 12 hour intervals)
- NIBP Max. Limit: 70 – 300 mmHg
- NIBP VENOUS STAT: 50 – 200 mmHg