V1410 / V1412 User Manual
86
Table for Labels According to the Site of Measurement
Label
Site
Measurement
ABP
Arterial Blood Pressure
Artery group throughout the whole body
(systemic)
CVP
Central Venous Pressure
Central vein / atrium group
ICP
Intracranial Pressure
Intracranial group
PAP
Pulmonary Arterial Pressure
Pulmonary artery group
Changing the IBP Label
1
To change the IBP label, press the Menu button one
time and select LABEL.
Mode
Setting
Range for Change
Description
SPEED
6.3, 12.5, 25, 50 mm/sec
Set the speed of IBP waveform.
LABEL
ABP, CVP, ICP, PAP
Set the level according to the
measurement site of BP. Change the alarm
setting range according to level.
UNIT
mmHg, kPa
Set the unit of IBP measurement.
* ZERO
ON
Adjust the zero point of the transducer.
SCALE
** OPTIMUM / FIXED
Set the wave scale.
TRACE
ON / OFF
Set the display of waveform.
* Adjust the zero point of the transducer: In order to obtain accurate results, make sure to adjust the zero
point of the transducer before measuring a patient.
** Fixed: Basic value according to the established label (ABP: 0–200, CVP: 0–50, ICP: 0–50, PAP: 0–50).
Optimum: Scale of waveform is automatically adjusted regardless of the level.