39. EtCO2 Measurement (Optional)
Microstream, mainstream, and sidestream CO
2
modules can be connected to the V1410 / V1412. The
microstream CO
2
module provides the function of EtCO2 and the function to measure respiratory CO
2
and respiration numbers using the small filter line based on a small lumen-type filter line. In the
microstream CO
2
type, the level is measured through the cannula (non-tubing type) attached to the nose
or a sampling line equipped in the respiratory circuit (tubing type).
WARNING
The EtCO2 measurement utilizes a 50 ml/min sampling rate. The monitor should not be
attached to patients that may become distressed with a 50 ml/min reduced inspiratory
volume.
The sampling outlet of the V1410 / V1412 should be attached to an active scavenging
system to evacuate the anesthetic gases discharged by the patient.
CAUTION
The gas inspiring vacuum pressure (negative pressure) of the gas elimination system should
not exceed the pump outlet standard of the V1410 / V1412, 1.0 mmHg. Excessive pressure
will display an OCCULSION message and it may damage the V1410 / V1412 pump. During
zeroing, the air elimination system should be operating.
Sampling lines and filters may become contaminated with patient bodily fluids. Use proper
protocols when handling such lines and accessories.
NOTE
The monitor zeros the CO
2
sensor during the power-up self test. Do not allow connection of
sample lines to the CO
2
source during the self test period.
Range of Respiration Measuring: 0 – 150 bpm
For the best accuracy, a warm-up of about 20 minutes is required.
It is optional to use 5% gas for calibration and the microstream filter line
EtCO2 Measurement Screen
No.
Description
No.
Description
1
CO
2
Waveform
5
FiCO2
2
CO
2
Waveform Speed
6
Respiration Measuring
3
CO
2
Waveform Size
7
EtCO2
4
Message Indicator