VTK-VS-BA-015e-A | User Manual MDS 3200+ Series
|
Error Messages
107
Limits:
NL
≤
80 % = 140 °C
NL > 80 % = 120 °C
(from 81 % – 100 % the temperature is variable at ca. 140 °C – 120 °C)
Once the temperature has fallen below 80 °C, the display changes from “MDV Temp. High
Please Wait” to “MDV Temp. High Press Enter”. After pressing the [Enter] key, you can again
trigger normally. During the display of “MDV Temp. High” the valve is closed.
If the error message happens directly after starting the MDC, pressing
[Enter]
will initiate a
re-start.
Error code display:
MDV Temp. High Please Wait
Error code status menu:
MDV Temp. High
Error handling:
•
The temperature of the valve is too high. System has to cool
off, then press
[enter]
. Maybe you have to lower the Needle
Lift and/or Frequency parameters, in order to prevent a re-
occurrence of the problem. In case of an air cooled valve,
you need to increase the air flow.
•
Switch OFF control unit, and inspect the connection. If a ca-
ble turns out to be damaged, it must be exchanged before
restarting.
Valve:
Valve is closed
9
Data Error Press Enter
During the check of RAM data, a problem arises. The message has to be acknowledged by
the
[enter]
key. The system reformats the EEPROM, replacing current values with the factory
settings.
Error code display:
Data Error Press Enter
Error code status menu:
Data Error
Error handling:
•
Press
[enter]
key to acknowledge error message.
•
Reenter the working configuration.
Valve:
Valve remains unchanged. If the error message appears during
the start of the MDC, the valve is open, otherwise it is closed.
10
Internal Error 701
A hardware failure of the valve control is probable
(e.g. short circuit in the connection of the
piezoelectric element)
. This problem is possible during startup of the control unit and also
with system in operation.
Error code display:
Internal Error 701
Error code status menu:
Internal Error 701
Error handling:
•
Switch OFF the control unit immediately. Inspect the actor
cable connection. Worn, kinked or charred cables must be
exchanged.
•
The valve must be returned to the manufacturer. The con-
trol unit has to be checked
(if that is not possible, it has to be
send as well)
.
Valve:
Valve is open
11
EEPROM not formatted Press Enter
Содержание MDS 3200+ Series
Страница 1: ......
Страница 117: ...VTK VS BA 015e A User Manual MDS 3200 Series Attachments 117 13 ATTACHMENTS 13 1 EC Declaration of Conformity...
Страница 118: ...118 VTK VS BA 015e A User Manual MDS 3200 Series Attachments 13 2 Dimensional Drawing MDC 3200...
Страница 119: ...VTK VS BA 015e A User Manual MDS 3200 Series Attachments 119 13 3 Dimensional Drawing MDV 3200A...
Страница 120: ...120 VTK VS BA 015e A User Manual MDS 3200 Series Attachments 13 4 F Dimensional Drawings MDV 3200F...
Страница 121: ...VTK VS BA 015e A User Manual MDS 3200 Series Attachments 121...
Страница 122: ...122 VTK VS BA 015e A User Manual MDS 3200 Series Attachments 13 5 Connection Diagram PLC Interface...