Warning Lamp & Door Faults
Red Electric Shock
Risk lamp lights
when door is
opened
Failed cut-off relay SR1 or
SR2
DO NOT USE until relay
replaced
Fuses for left side of
chamber blown
Check and replace fuse
Red Electric Shock
Risk lamp does not
light when tested
Faulty lamp
DO NOT USE until lamp
replaced
Faulty relay SR3 or SR4
DO NOT USE until relay
replaced
Faulty momentary switch
DO NOT USE until switch
replaced
Faulty door switch
DO NOT USE until door
switch replaced
Door does not
unlock
Furnace may have been
turned off during a test or
there may have been a
power failure during a test
Enter START PROCESS then
cancel the test.
THE RELEASE CODE '
728
' is
REQUIRED
Failure of interface board
Loose connection lead
Check all leads
Door does not lock
at start of test
The key has been used and
left in unlocked position
Set the manual override to
the locked position
Door switch failure
Interface board failure
Loose connection lead
Check all leads
E.
Chimney Emits Smoke
Some light smoke is
emitted in the
middle of the test
No fault
Dense smoke is
emitted
The after-burner elements
or control may have failed
Investigate or contact
Carbolite Gero
A fuse may have blown
Check the Supply fuses
87
27.0 Fault Finding
Содержание CARBOLITE GERO ABA 7/35B
Страница 11: ...2 2 Identifying the Parts 11 2 0 Supplied Item List ...
Страница 13: ...Fig 7 Front brick cover Fig 8 Calibration Plate Fig 9 Functional parts of the ABA 13 2 0 Supplied Item List ...
Страница 19: ...Fig 14 Schematic for Mains Electrical Connections 19 3 0 Installation ...
Страница 40: ...7 0 Running a Test Fig 46 Screen view Weight Entry Number Pad Fig 47 Screen view Successful Weight Entry 40 ...
Страница 57: ...Fig 61 Screen view Standard Recipe Fig 62 Calibration Recipe 57 14 0 Recipe Editor ...
Страница 70: ...20 0 Temperature Calibration Fig 72 Screen view Dual Point Calibration 70 ...
Страница 93: ...Notes Service Record Engineer Name Date Record of Work ...