Fig 43 - Screen-view: Weight Entry Screen
7.9
Connecting an Adam Equipment™ External Balance
Automatic weight entry requires an external balance to be connected to the 9 way
female D-type connector mounted at the rear of the furnace. To connect to an Adam
Equipment™ balance, use a 9 way, null modem, female to male cable. The required pin
connections are 2-3, 3-2, 5-5, all other pins are unused. The external balance type
must be set to Adam Equipment in the 'General Settings' screen (see Fig 45.).
The balance settings are shown below:
Baud Rate
9600
Parity
None
Stop Bits
1
Number of Bits
8
Refer to the Adam Equipment balance user manual for instructions on setting these
parameters.
7.10
Connecting an Ohaus Explorer™ External Balance
This section refers to the Ohaus Explorer balance with a membrane key pad and
monochrome display.
The external balance type must be set to Ohaus Explorer™ in the General Settings
screen (see Fig 45.).
37
7.0 Running a Test
Содержание CARBOLITE GERO ABA 7/35B
Страница 11: ...2 2 Identifying the Parts 11 2 0 Supplied Item List ...
Страница 13: ...Fig 7 Front brick cover Fig 8 Calibration Plate Fig 9 Functional parts of the ABA 13 2 0 Supplied Item List ...
Страница 19: ...Fig 14 Schematic for Mains Electrical Connections 19 3 0 Installation ...
Страница 40: ...7 0 Running a Test Fig 46 Screen view Weight Entry Number Pad Fig 47 Screen view Successful Weight Entry 40 ...
Страница 57: ...Fig 61 Screen view Standard Recipe Fig 62 Calibration Recipe 57 14 0 Recipe Editor ...
Страница 70: ...20 0 Temperature Calibration Fig 72 Screen view Dual Point Calibration 70 ...
Страница 93: ...Notes Service Record Engineer Name Date Record of Work ...