17.0 Internal Balance
17.0
Internal Balance Commands
17.1
Balance Command Screen
The balance command screen is accessible via:
Home Screen
→
Main Menu
→
Balance Menu
The screen is shown in Fig 65. The balance commands are:
l
Tare
l
Calibrate
l
Adjust Filter
If the balance is operating normally, then the balance status will display ‘Ready’. If
communication is lost between the balance and the controller, the balance status will
display ‘No reply to set up command’ and the cable between the controller and balance
should be checked.
Fig 67 - Screen-view: Balance Control Screen - Idle
17.2
Tare Command
Pressing the tare button will zero the balance by sending a tare command to the
balance. The balance status will display ‘Tare started…’ then ‘Tare complete’, for
approximately 2 seconds before displaying ‘Ready’ again. The weight reading should
read zero grams.