7.12
External Balance Configuration
The external balance type must be configured in the 'General Settings' screen (see Fig
45.).
This screen is accessed via:
Home Screen
→
Main Menu
→
System Configuration
→
General Settings
Fig 45 - Screen-view: External Balance Type Configuration
When the Enter Weights button is pressed, the controller will attempt to communicate
with an external balance. If it successful then external balance status on the Weight
Entry screen will read ‘Ready’ (see Fig 43.). If it fails then ‘No reply to weight request’
will be displayed. To retry the connection, return to the Home screen and then press
‘Enter Weights’ again. If a successful connection is made then ‘Ready’ will be displayed.
7.13
Entering Weights
To enter weights just two weights are required, the third weight will be calculated.
There is no specific order required. If using automatic weight entry, just place the item
to be weighed on the external balance and press either ‘Get Tray Weight’, ‘Get
Combined Weight’ or ‘Get Sample Weight’ as appropriate. The relevant weight will
appear opposite the button.
To manually enter weights, press the white box for the required weight and the number
pad will be displayed as shown in Fig 46.
Note; the top of the number pad indicates the item requiring a weight with the units
and the maximum and minimum weight range.
Weights can be re-entered at any time by pressing the white box again. Also the ‘Clear
Weights’ button can be used at any time. Once the weights have been entered the
‘back’ button changes from a yellow arrow to a green tick (compare Fig 43 with Fig 47).
Pressing the green tick causes the Home screen to be displayed, with the newly entered
weights shown in the top right quarter of the screen. If the ‘Enter Weights’ button is
pressed again then the current weight readings will be displayed on the Weight Entry
screen, allowing the operator to amend just one or more values either manually or
automatically.
39
7.0 Running a Test
Содержание CARBOLITE GERO ABA 7/35B
Страница 11: ...2 2 Identifying the Parts 11 2 0 Supplied Item List ...
Страница 13: ...Fig 7 Front brick cover Fig 8 Calibration Plate Fig 9 Functional parts of the ABA 13 2 0 Supplied Item List ...
Страница 19: ...Fig 14 Schematic for Mains Electrical Connections 19 3 0 Installation ...
Страница 40: ...7 0 Running a Test Fig 46 Screen view Weight Entry Number Pad Fig 47 Screen view Successful Weight Entry 40 ...
Страница 57: ...Fig 61 Screen view Standard Recipe Fig 62 Calibration Recipe 57 14 0 Recipe Editor ...
Страница 70: ...20 0 Temperature Calibration Fig 72 Screen view Dual Point Calibration 70 ...
Страница 93: ...Notes Service Record Engineer Name Date Record of Work ...