21.0 Maintenance
Fig 73 - Screen-view: Hours of Use
Fig 74 - Screen-view: Cleaning Warning Message
21.5
Cleaning the Plenum Chamber & Fan Impeller
Remove the upper side access panel and the cover of the plenum chamber (see Fig 75.
and Fig 76.). Clean out all accumulated soot and debris using a vacuum cleaner. Also
clean any accumulated soot and debris from the fan impeller located at the back of the
plenum chamber. If the accumulation of soot on the impeller is significant, it may be
necessary to remove the motor/ impeller assembly from the back of the plenum
chamber to give good access for cleaning.
74
Содержание CARBOLITE GERO ABA 7/35B
Страница 11: ...2 2 Identifying the Parts 11 2 0 Supplied Item List ...
Страница 13: ...Fig 7 Front brick cover Fig 8 Calibration Plate Fig 9 Functional parts of the ABA 13 2 0 Supplied Item List ...
Страница 19: ...Fig 14 Schematic for Mains Electrical Connections 19 3 0 Installation ...
Страница 40: ...7 0 Running a Test Fig 46 Screen view Weight Entry Number Pad Fig 47 Screen view Successful Weight Entry 40 ...
Страница 57: ...Fig 61 Screen view Standard Recipe Fig 62 Calibration Recipe 57 14 0 Recipe Editor ...
Страница 70: ...20 0 Temperature Calibration Fig 72 Screen view Dual Point Calibration 70 ...
Страница 93: ...Notes Service Record Engineer Name Date Record of Work ...