4.0 Safety Features
The user is responsible for maintaining a clean plenum chamber and also for regular
cleaning of the chimney and the external flue or ducting system.
4.4
Door Lock Indication 'Lamps' on Touchscreen
Fig 32 - Screen-view: Top Status Bar
A ‘door locked lamp’ is shown on the top status bar on the user interface (see Fig 32.). If
the door is locked then it is illuminated in orange, otherwise it is grey. The user should
always check that the ‘door locked lamp’ is lit when a test is started.
4.5
Maintenance or Dismantling
NOTE: This product contains Refractory Ceramic Fibre (better described as
Alumino Silicate Wool) for precautions and advice in handling this material
see the "Safety Warning - Refractory Fibre Insulation".
The user should disconnect the equipment from the electrical supply before removing
panels to access the electrical connections and control equipment. Caution : Double
pole/ neutral fusing may be used in this product.