8 Diagnosis, asset management and service
VEGABAR 86 • Secondary Device for electronic differential pressure
48048-EN-220817
1
2
3
6
4
5
Fig. 23: VEGABAR 86 in IP68 version, 25 bar and lateral cable outlet, external
housing
1 Process module
2 Plug connector
3 Cable assembly
4 Connection cable
5 External housing
3. Loosen the plug connector
4. Mount the new process module on the measuring point
5. Plug the connector back in
6. Mount the cable assembly on the process module and turn it to
the desired position
7. Tighten the fixing screw with the hexagon key wrench
The exchange is finished.
8.5 How to proceed if a repair is necessary
You can find an instrument return form as well as detailed information
about the procedure in the download area of our homepage. By doing
this you help us carry out the repair quickly and without having to call
back for needed information.
Proceed as follows in case of repair:
•
Print and fill out one form per instrument
•
Clean the instrument and pack it damage-proof
•
Attach the completed form and, if need be, also a safety data
sheet outside on the packaging
•
Ask the agency serving you to get the address for the return ship-
ment. You can find the agency on our homepage.