vNet Power over Ethernet Device User Guide
15
2
In the NetBIOS name field you can enter up to 15
characters (using letters, numbers, and underscores). Do
not assign the same name to another device.
3
Click
Apply
.
4
A message will appear stating that the vNet device
needs to be restarted. Click on the
here
link to restart
the device.
Changing Ethernet Settings
Depending on the network hub or switch the vNet device is
connected to, you may want to change the Ethernet mode
from the default setting, half-duplex, to full-duplex.
To change the Ethernet settings for your vNet device:
1
From the vNet Configuration and Management interface,
choose
Configuration | Network | Ethernet Settings
.
2
Choose the desired Duplex Mode (Half or Full).
3
Click
Apply
.
4
A message will appear stating that the vNet device
needs to be restarted. Click on the
here
link to restart
the device.
Note:
The speed setting is fixed at 10Mbit and cannot
be changed.
Editing Serial Port Settings
Use the Serial Port Settings screen to set the TCP port
numbers that the vNet device will use for serial
communication. The vNet device communicates with the
Veriteq data logger through a serial port.
The serial port is accessed by the client PC using serial port
redirector (virtual COM port) software that communicates
over the Ethernet connection using TCP ports. The vNet
device needs to be configured to use the same TCP port as
the serial port redirector software.