9. Technical Data
9.1 Specifications
Table 6 Input and Output
Property
Specification
Digital output
RS-485 Modbus RTU
Relays
2 configurable relays (VAC/VDC)
Device maximum specification (resistive load):
• Max. switching power 30 W / 37.5 VA
UL-rated maximum specification (resistive load):
• AC: max. 28 V / 0.5 A
• DC: max. 40 V / 0.24 A
• Up to 30 VDC:
• max. switching current 1 A
• max. switching power 30 W
Range 15 ... 30 VDC (20 ... 22 VAC)
Maximum current
Transmitter and connected probe max. 1 A
Power consumption
Transmitter max. 3 W (+ connected probe, varies
depending on probe type)
Probe connector
M12/5 connector for probe or probe cable connection
(Vaisala Indigo-compatible probes)
Cable feed throughs
2 options: rubber lead-through on the bottom of the
transmitter, and opening with a seal at the back of the
transmitter
Screw terminal wire size
0.2 mm
2
... 1.5 mm
2
1)
Using a power supply with overload protection is recommended for electrical safety.
Table 7 General
Display
3.5” TFT LCD color display
Configuration interface
Browser-based wireless configuration interface (IEEE
802.11 b/g/n WLAN)
Wireless configuration interface browser support
• Microsoft Internet Explorer: version 11.0 onward
• Google Chrome: version 57 onward
• Mozilla Firefox: version 50 onward
• Apple Safari: version 10 onward
Chapter 9 – Technical Data
61
Содержание Indigo 202
Страница 1: ...M211966EN A User Guide Indigo 202 Digital Transmitter Indigo 202...
Страница 22: ...Indigo 202 User Guide M211966EN A 20...
Страница 62: ...Indigo 202 User Guide M211966EN A 60...
Страница 70: ...Indigo 202 User Guide M211966EN A 68...
Страница 74: ...Indigo 202 User Guide M211966EN A 72...
Страница 75: ......
Страница 76: ...www vaisala com...