To make a 2-point adjustment (GMP252 ppmCO
2
example):
1. Connect to the wireless configuration interface and open the
Calibration
menu.
2. Start the calibration mode with the
Start calibration
button.
If you cannot enter configurations after selecting
Start calibration
, check that the
calibration PIN code is in place in the
Settings > Probe
menu.
3. If you need to set environmental compensations, enable and set the required
compensations on the
Configuration
tab.
Note that the environmental compensations you set on the
Configuration
tab and
the compensations you set in the
Settings > Probe
menu are interconnected: the
configuration set in either menu is applied to both.
4. Remove any possible previous adjustments by restoring the factory adjustment: select
Restore factory adjustment
for each parameter you are adjusting.
5. Enter the calibration date and calibration information into the corresponding text fields.
6. Place the probe in the first reference environment (adjustment point 1) and wait until
the measurement has stabilized. You can follow the stabilization from the
Measurements
tab.
7. Enter the value of the first reference (for example,
0
if calibrating with a 0 ppmCO
2
reference gas) into the
Reference value, point 1
field.
Chapter 7 – Calibration and Adjustment
53
Содержание Indigo 202
Страница 1: ...M211966EN A User Guide Indigo 202 Digital Transmitter Indigo 202...
Страница 22: ...Indigo 202 User Guide M211966EN A 20...
Страница 62: ...Indigo 202 User Guide M211966EN A 60...
Страница 70: ...Indigo 202 User Guide M211966EN A 68...
Страница 74: ...Indigo 202 User Guide M211966EN A 72...
Страница 75: ......
Страница 76: ...www vaisala com...