8. After you enter the reference value, the value of the
Measured value, point 1
field
updates automatically.
9. Place the probe in the second reference environment (adjustment point 2) and wait
until the measurement has stabilized.
10. Enter the value of the second reference (for example,
2000
if calibrating with a 2000
ppmCO
2
reference gas) into the
Reference value, point 2
field.
11. When both reference points have been entered, select
Store adjustment
to save the
adjustment.
12. Close the calibration mode with the
Stop calibration
button.
13. To check that the adjustment was carried out correctly, review the information in the
Adjustment data
fields at the bottom of the view.
Indigo 202 User Guide
M211966EN-A
54