Compensation Power-up Defaults (GMP251 Example)
The power-up default values are stored to the probe memory and remain in use also after
you disconnect or reset the probe.
Note that the environmental compensations you set in the
Settings > Probe
menu and the
compensations you set in the
Calibration
menu are interconnected: the configuration set in
either menu is applied to both.
More Information
‣
Environmental Compensation (page 48)
Indigo 202 User Guide
M211966EN-A
38