Table 15 Modbus Status Registers (Read-Only)
Address
(Hexadecimal)
Register Description
Data Format
Notes
E000
hex
Status
16-bit
0 = Status OK
1 = Info
2 = Warning
3 = Error
E001
hex
Notification and error
bits
16-bit
Provides additional details on the
high-level status reported in
register E000
hex
E011
hex
Relay A status
enum
0 = Relay is not active
1 = Relay is activated
E012
hex
Relay B status
enum
0 = Relay is not active
1 = Relay is activated
A.2.1 Notification and Error Bits
The status register (E000
hex
) can report that one of the following notification types is
present:
0
No issues
1
Info
2
Warning
3
Error
Additional details about the status reported in the register E000
hex
can be provided in the
16-bit register E001
hex
, as described in the following table.
Table 16 Bits in Register E001
hex
Bit in E001
hex
Meaning
Notes
0 (0000
hex
)
No errors
1 (0001
hex
)
WLAN on
Wireless configuration interface is
activated and can be connected to.
2 (0002
hex
)
Probe error
Connect to Indigo's wireless
configuration interface and review
the error information in the
Status
menu. Check display messages.
4 (0004
hex
)
Low supply voltage
The supply voltage is not within
the specified range (15 ... 30 VDC
(20 ... 22 VAC)
8 (0008
hex
)
WLAN error
See the WLAN troubleshooting
instructions.
Indigo 202 User Guide
M211966EN-A
66
Содержание Indigo 202
Страница 1: ...M211966EN A User Guide Indigo 202 Digital Transmitter Indigo 202...
Страница 22: ...Indigo 202 User Guide M211966EN A 20...
Страница 62: ...Indigo 202 User Guide M211966EN A 60...
Страница 70: ...Indigo 202 User Guide M211966EN A 68...
Страница 74: ...Indigo 202 User Guide M211966EN A 72...
Страница 75: ......
Страница 76: ...www vaisala com...