7.2 Environmental Compensation
When making adjustments, you can enter information about the environment in which you
perform the adjustment. This allows compensating for environmental factors that have an
effect on the measurement (for example, temperature, pressure, or background gases). The
selection of environmental compensations available in the Indigo calibration settings
depends on the features and configuration of the connected probe.
The environmental compensation selections are available on the
Configuration
tab of the
Calibration
menu. To use environmental compensations, first enable the compensation from
the
Measurement
selections, and then enter the compensation in the
Compensation
setpoints
fields. You can also set the power-up default compensations that remain in use
also after probe reset.
Measurement
Figure 23 Measurement Selections, CO
2
Probe Example
Filtering factor
Defines how much past measurements affect the output
(measurement averaging over time). For details, see probe
documentation.
Temperature
compensation mode
Select whether the probe sensor's measurement or a manually
entered setpoint is used to set the temperature compensation.
Pressure/humidity/oxygen
compensation on/off
Enable or disable the environmental compensations.
Indigo 202 User Guide
M211966EN-A
48
Содержание Indigo 202
Страница 1: ...M211966EN A User Guide Indigo 202 Digital Transmitter Indigo 202...
Страница 22: ...Indigo 202 User Guide M211966EN A 20...
Страница 62: ...Indigo 202 User Guide M211966EN A 60...
Страница 70: ...Indigo 202 User Guide M211966EN A 68...
Страница 74: ...Indigo 202 User Guide M211966EN A 72...
Страница 75: ......
Страница 76: ...www vaisala com...