![Vaisala DMP248 Скачать руководство пользователя страница 72](http://html1.mh-extra.com/html/vaisala/dmp248/dmp248_operating-manual_3216965072.webp)
OPERATING MANUAL _______________________________________________________________
68_____________________________________________________________________M210492
EN
-
A
ERRS <cr>
If there are no error messages, only a prompt is displayed:
>ERRS <cr>
>
When errors have occurred, the transmitter outputs the error code (see
Appendix 5 for all error messages):
>ERRS <cr>
E40 f ( all ) out of range
>
Temperature channel adjustment with Pt 100
simulators
Switch the power off and disconnect the wires to the Pt 100 sensor
from solder lugs TP5, TP6 and TP7.
TP6
TP5
TP7
X88
F
IGURE
7-1
Location of solder lugs TP5, TP6 and TP7 and
connector X88
Connect a Pt 100 simulator to connector X88 and set it at the lowest
temperature to be calibrated.
X88
Pt 100
F
IGURE
7-2
Connecting the Pt 100 simulator to connector X88
Switch the power on.
All manuals and user guides at all-guides.com
Содержание DMP248
Страница 6: ...All manuals and user guides at all guides com a l l g u i d e s c o m...
Страница 28: ...All manuals and user guides at all guides com...
Страница 36: ...All manuals and user guides at all guides com a l l g u i d e s c o m...
Страница 70: ...All manuals and user guides at all guides com...
Страница 82: ...All manuals and user guides at all guides com...
Страница 84: ...All manuals and user guides at all guides com...
Страница 108: ...All manuals and user guides at all guides com...
Страница 133: ...APPENDIX 6 All manuals and user guides at all guides com...
Страница 135: ...All manuals and user guides at all guides com...
Страница 140: ...All manuals and user guides at all guides com...