Chapter
6
■
Output data select
≪
Setting value
≫
Hold Sync., Hold Not Sync., Zero Scale, Full Scale
Hold Sync.:
Output is produced in synchronization with the indicated value.
Hold Not Sync.:
Outputs value without hold even when the indicated value is on hold.
Zero Scale:
Output is fixed at zero output (0V).
Full Scale:
Output is fixed at full scale (10V).
* Set torque, rotation speed and power, respectively.
How to set
How to output adjust
Select each fixed output in D/A output select, and press the
button, and then use the
simulated trimmer appearing on the adjustment screen to adjust the output.
While monitoring the output value, make fine adjustment of the output with the UP/DOWN
button of the simulated trimmer, and determine by pressing the
button, so
that the trimmer position is entered.
* Pressing the simulated trimmer
buttons continuously performs continuous
operation.
* The trimmer adjusting ranges are as follows: voltage output; about
±
2.0V.
* The [Clear] button resets the Simulated trimmer position to the trimmer center point (0%).
SET.
Option Setting (D/A Output)
Page 1
→
System Setting
→
→
Clear button
Simulated trimmer
Simulated trimmer
adjustment
(The trimmer moves to
the directly-pressed place.)
Output channel during
Scale during adjustment
UP
・
DOWN button
position display
adjustment