4
1
OUTLINE
4
O
U
TL
IN
E
Chapter
1
DC power input terminal block
Connect the DC power cord. The power supply voltage is 24V DC (±15%).
Adaptable crimp terminal [TMEV1.25-3S]
Frame ground
Please ground the frame ground terminal to prevent failures due to static electricity.
(The frame and the frame ground terminal are conducted.)
It may be better to remove depending on the environment of the installation location.
Protective ground
Be sure to ground the protective ground terminal to prevent electric shocks.
Voltage output terminal
The approximate input voltage from UTM
Ⅱ
can be confirmed (approx. 1V per sensor input voltage
of 1V, with load resistance of 2kΩ or more).
* For connection, see "
USB connector
This connector is used to connect the product with a computer.
* For connection, see "
USB interface connection" on page 11.
Optional slot
D/A converter voltage output (3ch) [D3V] can be installed.