18
4
CALIBRATION
18
C
A
LIB
R
A
TIO
N
Chapter
4
4-3.
Calibration Value Selection
(optional if there is no change)
By storing up to two calibration values in the memory, the desired calibration value can be called to
switch the indicated value. Setting values that can be switched are as follows:
≪
Setting value
≫
0, 1, EXT*
(* “EXT” stands for “External Signal”: OFF=0, ON=1)
How to set
1.
Press the [SET.] button. (see P.17)
2.
Press the [Exp. Setting] button.
3.
Press the [Exp. Calibration] button.
■
Calibration value selection by touch panel
Select "0 - 1" by pressing the [
∨∧
] button on the Exp. Calibration screen.
When not using, set "0".
■
Calibration value selection by external signal input
With this function, two types of calibration values can be selected with external selector signals
CAL (when the calibration value selection setting is external).
Set "EXT" 0 by pressing the [
∨ ∧
] button on the Exp. Calibration screen. According to the input
conditions of external signal inputs CAL, the display changes as EXT0 to EXT1. Normally, when
there is no input to CAL (the terminals are open), calibration value 0 is selected.
SET.
Exp. Setting
Exp. Calibration
→
→
Page 1
→
Calibration mode setting
Unit (torque)
Unit (power)
Zero calibration (torque)
Increment (power)
Actual load calibration (torque) Decimal place (power)
Equivalent input calibration
(torque)
D/A setting (torque)
Zero scale
Full scale
Increment (torque)
Decimal place (torque)
Digital zero limit
D/A setting (speed)
Zero scale
Full scale
Decimal place (speed)
Increment (speed)
Pulse input
D/A setting (power)
Zero scale
Full scale
Pulse rate
Minimum input rotation speed
Rotation stop mode
Содержание TM700
Страница 1: ...20MAY2021REV 1 08 TORQUE MONITOR TM700 OPERATION MANUAL ...
Страница 91: ......