* Command definition
3. Normal [operation command]
Setting value reading
Function
Header
Protocol
Main
Sub
Various
settings
W
-
Operation directive
Function
Header
Protocol
Main
Sub
Zero
calibration
C
Z
Equivalent
input
calibration
C
S
Digital zero
C
G
Digital zero
reset
C
H
W
W
CR
Delimiter
Host
TM700
Setting value (Sign, 5-digit, no decimal place)
(A setting value with a decimal place results
in the numerical value excluding the decimal place.)
Command
Command
* For command numbers,
see "8-4.Setting Item List" on page 75.
* *
Command number (×1)
Command number (×10)
- "Initialize" and "password" cannot be either read or written.
- When the protect of the setting value you are going to write is ON, it cannot be
written. Turn OFF the protect.
- If an output value is written in command 63 (zero calibration), simulated zero
calibration is executed with the written value. Also, if an output value is written
in command 64 (equivalent input calibration), equivalent input calibration is
executed.
Point
C
CR
Host
TM700
Z
C
CR
Host
TM700
S
* Set the rated capacity value with command 65 before sending
a command for actual load calibration.
C
CR
Host
TM700
G
C
CR
Host
TM700
H