Hysteresis
The hysteresis value may be determined so as to allow a margin for timing the turning off of the HI/
LO limit comparison. Normally, it is turned on when the indicated value exceeds the HI limit and is
turned off when the indicated value falls below it. However, by setting the hysteresis, it is turned off
when the indicated value falls below the HI limit further lowered by the hysteresis value.
This function is effective to prevent chattering in such a case where signals fluctuate (vibrate)
subtly.
≪
Setting value
≫
0 to 9999
<
HI/LO output conditions
>
- HI
ON conditions
Indicated value
>
HI limit setting value
OFF conditions Indicated value
≦
(HI limit setting value
-
Hysteresis setting value)
- LO
ON conditions
Indicated value
<
LO limit setting value
OFF conditions Indicated value
≧
(LO limit setting value
-
Hysteresis setting value)
How to set
- Hysteresis operation
SET.
Comp. Setting
Page 1/Page 3
→
→
Meas. Setting
→
HI limit
LO limit
+
0
Hysteresis range
Time
HI output
HI status display
LO output
LO status display
OK status display
Output ON
Status ON
Output OFF
Status OFF
Status ON
Status ON
Output ON
Status ON
Output ON
Status ON
-
Output OFF
Status OFF
Output OFF
Status OFF
Indicated value
OK output
(Exsample: HI output and LO output and OK output)
Torque and rotation speed can be set to HI/LO limit comparison, respectively.
However, this does not apply to alarm HI and LO limits.