53
5
SETTING OF FUNCTIONS
53
SE
TT
IN
G
O
F
FU
N
C
TI
O
N
S
Chapter
5
5-18.
USB interface
The USB is an interface to read the indicated value of the TM700 and to write parameters into the
TM700. This interface is convenient to process controls, totals, records, etc., by connecting the
TM700 to PC. Furthermore, reading and writing setting values, reading graph data and capturing
screens are possible using a dedicated PC application.
■
Communication specifications
Communication standard:
Compliant with USB Ver.2.0, Full speed (12Mbps)
Class:
Communication device class
Baud rate:
9600, 19200, 38400, 57600, 115.2k, 230.4k bps selectable
Bit configuration:
Start bit:
1bit
Character length:
7 or 8 bit selectable
Stop bit:
1 or 2 bit selectable
Parity:
none, odd or even selectable
Delimiter:
CR, CR+LF selectable
Code:
ASCII
Connector:
mini-B TYPE
■
PC hardware requirement (TM700 PC application software)
OS:
Windows7 Home Premium/Professional/Ultimate 32/64bit
Display:
1024×768 pixels
(XGA) or more
USB port:
One empty port
USB driver:
Virtual COM Port(VCP) Drivers by FTDI Ltd.
■
Installation of a USB driver
Please install a USB driver for connecting with USB of the TM700.
Please see the homepage of FTDI for details and perform download and installation of the driver
according to a procedure.
* If the PC is already connected to the network, the driver may be automatically installed via the
network when the v is connected to the PC.
Guide:
http://www.ftdichip.com/Support/Documents/InstallGuides.htm
Driver:
http://www.ftdichip.com/Drivers/VCP.htm
■
Connection of USB
Connects a USB cable.
The USB connector of the TM700 is mini-B TYPE.
■
The check of a virtual COM port
Please check the virtual COM port number which the TM700 is connected from the device manager
of PC.