CWD2005 DPC
Connections, transport, setting up
chapter — page:
5—15
5.4.8 Calibration gas
WARNING
Only qualified trained staff must install the gas connections!
If no pressure reducers are installed, the operator must make sure that escaping
calibration gas is emitted to a safe environment!
NOTE
•
The connecting parts need to be clean and free of residue. Impurities can en-
ter the combustion calorimeter and cause incorrect measurements and/or
damage.
•
Calibration gas connection / clamping ring screw connection on the side of
the combustion calorimeter.
•
The inlet pressure for the gas connections must not exceed the values speci-
fied in the technical data of the combustion calorimeter.
•
Calibration gas must be free from impurities and condensate.
•
The quality of the calibration gas should be similar to the process gas quality.
•
Each connection needs to be carefully checked for leaks. If there are any
leaks, the system will draw air, and the measurements will be incorrect.
•
Do not use sealing agents to seal the gas connections. Sealing agents can
distort measurements.
•
Only use suitable pipes.
•
For device variants 2 or 3 measuring ranges, up to 2 calibration gases can be
required.
•
Software configuration of the calibration gas.
(
chapter 7.3 Present displays)
Содержание CWD2005 DPC
Страница 1: ...Operating instructions Combustion calorimeter CWD2005 DPC Mai 2017 V2 01 ...
Страница 12: ...CWD2005 DPC chapter page 2 2 ...
Страница 28: ...CWD2005 DPC chapter page 4 6 ...
Страница 114: ...CWD2005 DPC chapter page 9 2 ...
Страница 116: ...CWD2005 DPC chapter page 10 2 ...
Страница 119: ...CWD2005 DPC Annex chapter page 11 5 ...
Страница 120: ...CWD2005 DPC Annex February 2015 V2 00 chapter page 11 6 ...