CWD2005 DPC
chapter — page:
7—48
7.6.1.1.2.1 Main menu - Options - Calibration - Configuration of calibration gas
The calibration gas values are entered as Wobbe-index and relative density. These
values are calculated from the gas components of the calibration gas if the calibra-
tion gas producer only specified %-numbers of the gas components and not dv and
heating value. Take into account of the calorific value or the heating value is to be
used for the calculation.
The literature values of the components refer to dry or humid gas. The temperature
reference point (0°C or 15°C) also results in a difference in reference value. The
producer always bases the specification on 0°C and 1013 mbar with dry gas.
Содержание CWD2005 DPC
Страница 1: ...Operating instructions Combustion calorimeter CWD2005 DPC Mai 2017 V2 01 ...
Страница 12: ...CWD2005 DPC chapter page 2 2 ...
Страница 28: ...CWD2005 DPC chapter page 4 6 ...
Страница 114: ...CWD2005 DPC chapter page 9 2 ...
Страница 116: ...CWD2005 DPC chapter page 10 2 ...
Страница 119: ...CWD2005 DPC Annex chapter page 11 5 ...
Страница 120: ...CWD2005 DPC Annex February 2015 V2 00 chapter page 11 6 ...