CWD2005 DPC
Maintenance and service
chapter — page:
8—7
NOTE
The following overview only describes the checks/work in regard to the combus-
tion calorimeter!
Inspection
Interval
(recommended)
Weekly inspection
Check air filter for dirt / clogging.
weekly
Calibrate (depending on the requirements for precision, shorten the
intervals).
weekly
Monthly check
Check neoprene tubes in the combustion calorimeter for porosity.
monthly
Half-yearly inspection
Check the tube connection to the density measuring cell / sensors
for porosity.
every 6 months
Check the membrane at the gas pressure controller for porosity or
insufficient elasticity.
every 6 months
Remove
2
/clean
3
thermal battery and heat exchanger.
every 6 months
Replace the seal at the heat exchanger.
every 6 months
Check fan.
every 6 months
Check inlet filter (inlet ambient air, filter mat, ventilator).
every 6 months
Clean
4
gas and air nozzles. If necessary, replace O-rings.
every 6 months
Yearly inspection
Replace air filter.
yearly
Replace neoprene tubes in the combustion calorimeter.
yearly
Clean the complete system.
yearly
2
On the side of the combustion calorimeter, at the thermal battery and in the thermal body, flow mem-
bers are attached. These members must not be twisted when taking out the thermal battery. Pull out
the thermal battery as far as possible and then flip it sideways.
3
Clean the heat exchanger with water and then dry the cleaned components thoroughly.
4
Clean the nozzles with a solvent (brake cleaner, degreasing spray).
Содержание CWD2005 DPC
Страница 1: ...Operating instructions Combustion calorimeter CWD2005 DPC Mai 2017 V2 01 ...
Страница 12: ...CWD2005 DPC chapter page 2 2 ...
Страница 28: ...CWD2005 DPC chapter page 4 6 ...
Страница 114: ...CWD2005 DPC chapter page 9 2 ...
Страница 116: ...CWD2005 DPC chapter page 10 2 ...
Страница 119: ...CWD2005 DPC Annex chapter page 11 5 ...
Страница 120: ...CWD2005 DPC Annex February 2015 V2 00 chapter page 11 6 ...