![Union CWD2005 DPC Скачать руководство пользователя страница 42](http://html.mh-extra.com/html/union/cwd2005-dpc/cwd2005-dpc_operating-instructions-manual_842961042.webp)
CWD2005 DPC
chapter — page:
5—14
5.4.7 Carrier gas supply
NOTE
•
Combustion calorimeters can later be refit to carrier gas supply. Contact the
manufacturer.
•
Process gases without stable combustion properties can be mixed with a car-
rier gas.
•
Non-flammable gases are automatically mixed with flammable or not flamma-
ble carrier gases by the combustion calorimeter. The carrier gas will be reck-
oned up in the combustion calorimeter.
•
Carrier gases maintain the combustion and do not increase the calorific value
of the process gas.
carrier gas
consumption
Ø nozzle
methane
10 l/h
0,40 mm
propane
6,0 l/h
0,40 mm
hydrogen
51,5 l/h
0,30 mm