![Union CWD2005 DPC Скачать руководство пользователя страница 86](http://html.mh-extra.com/html/union/cwd2005-dpc/cwd2005-dpc_operating-instructions-manual_842961086.webp)
CWD2005 DPC
chapter — page:
7—58
7.6.1.1.3.10
Main menu - Options - System - Settings - Hardware 1
This menu should only be modified after consulting the manufacturer.
It controls the limits of measuring devices over two or three ranges. The overlap is
optimised. For this purpose you need to specify functional parameters. The time
constants of the burners are specified.
The calibration of the values of the carrier gas is entered and the time of connection
is determined. In case of changing the carrier gas, the manufacturer can provide
the new data. They are entered in the screen mask shown below.
All values are tested according to customer's specifications before delivery.
The screen shows all possible configurations.
Range No.
indicates the number of installed measuring ranges. A maximum of 3
measuring ranges are possible. In this example, 3 measuring ranges are config-
ured.
Dia. WJet.
indicates the nozzle diameter in mm of the measuring range nozzle.
This specification is information only and is meant to help later in calculating the
gas volumes.
Содержание CWD2005 DPC
Страница 1: ...Operating instructions Combustion calorimeter CWD2005 DPC Mai 2017 V2 01 ...
Страница 12: ...CWD2005 DPC chapter page 2 2 ...
Страница 28: ...CWD2005 DPC chapter page 4 6 ...
Страница 114: ...CWD2005 DPC chapter page 9 2 ...
Страница 116: ...CWD2005 DPC chapter page 10 2 ...
Страница 119: ...CWD2005 DPC Annex chapter page 11 5 ...
Страница 120: ...CWD2005 DPC Annex February 2015 V2 00 chapter page 11 6 ...