-10-
4.Installation
4.1 Installation of a pump
MBS -053/MBS-053-01 becomes design to use in an inline.
Fix flange of the outlet side, and install a pump in this flange.
Note
Install a pump horizontally. Horizontal admissible value of a pump longer
direction is ± 0.5 degrees.
Must not install a pump diagonally.
Be inverted,
and must not install it.
There is the case that lubricating failure and inflow of
lubricating oil to rotor cylinder come unless a pump is installed horizontally.
4.2 Environmental conditions for storage, installation and operation
Since this pump is precisely engineered, ensure that the following conditions
be satisfied during storage, installation and operation.
①
Ambient temperature, relative humidity: 0°C-40°C, 85% RH or less
②
Height above sea level during storage and installation: 1,000 m or less
③
Minimum required distances from the wall, 100mm
④
Other conditions for storage and operation
a.
Free from corrosive and explosive gases
b.
No condensation
c.
Dust-free environment
d.
Indoors
e.
Do not place pumps on top of each other or place a pump on its side.
f.
Not subject to direct sunlight
g.
Far from heat sources
h.
When you keep it for a long period of time, put pump oil into a pump
and block up inlet port and outlet port.
i.
Don't keep it, where moisture is attracted.
Warning
Do not use the pump in an explosive atmosphere. Failure to do so will result in
injury or fire.
Note
Minimum required distances from the wall, 100mm.