- 13 -
4. Operation
4.1 Cautions for operation
There is a risk of explosion. Never block the outlet or operate the pump with
equipment mounted at the outlet side which blocks the passage of gas.
Otherwise, the pump internal pressure increases causing the pump to explode,
the oil level gauge to protrude or the motor to be overloaded.
This pump is not resistant to pressure. The internal pump pressure is limited
to 0.03 MPa (gauge pressure).
Do not stop and restart this equipment repeatedly. When restarting this
equipment, make sure that the pump (motor) has stopped completely before
turning on the power. If the pump (motor) has not stopped completely and the
equipment is restarted, the pump’s (motor) current rises and either the
protection device is activated or the motor may become damaged.
① In the process of manufacturing semiconductors, pump oil may deteriorate
over a very short period of time. It is recommended that the pump oil should
be replaced within 10 days after starting use of the pump, and the
replacement frequency of the pump oil should be decided based on the
contamination level of the pump oil.
② If the pump evacuate a lot of moisture, replace the oil frequently. If
the pump is used with gas which contains a lot of moisture, water
absorption expands the vanes of the pump, the lubricity of the pump oil
deteriorates and corrosion of the pump’s components advance, causing the
pump to malfunction.
③ If chemicals including acid has been evacuated, the pump may become rusty
while it is not being operated (i.e. overnight), making operation
impossible. If such chemicals are evacuated, replace the pump oil
immediately.
④ Solvents which deteriorate the lubricity of the pump oil will cause
scoring, etc. If such a solvent is evacuated, replace the oil.
⑤ If operation is performed continuously at a high evacuation pressure of
10 kPa or more, a large amount of pump oil is consumed, causing a shortage
of oil and insufficient lubrication of the pump. If such a condition
continues, components will rapidly wear and become scored. Avoid
continuous operation at a high evacuation pressure as much as possible
and, without fail, add pump oil.
⑥ Do not block the flow of air to the motor fan as the temperature of the
motor and pump will increase.
Warning
Note
Caution
Содержание GLD-137CC
Страница 18: ...4 2 2 Dimensional drawing Fig 1 Dimensional drawing of GLD 137CC oil sealed rotary vacuum pump...
Страница 19: ...5 Fig 2 Dimensional drawing of GLD 202BB oil sealed rotary vacuum pump...
Страница 33: ...19 Fig 10 Pumping speed curve...
Страница 44: ...30 8 2 Disassembly drawing Fig 12 Disassembly drawing of GLD 137CC oil sealed rotary vacuum pump...
Страница 45: ...31 Fig 13 Disassembly drawing of GLD 202BB oil sealed rotary vacuum pump...