38
YK18-0012-AI-101-02
6.2 Alarm state and the responding action
Display
Alarm name
State
P1 ALARM
Abnormality of the device
specified in P1
Cause
Responding method
Reference
An abnormal signal is received
from the device specified in P1.
Handle the matter according to the instruction manual
of the device.
Display
Alarm name
State
P2 ALARM
Abnormality of the device
specified in P1
Display
Alarm name
State
TMP ALARM
TMP detects an abnormality.
Cause
Responding method
Reference
Overload on TMP
Use the delay timer to delay the TMP startup time and
operate the machine.
2.4.5
6.1
Disconnection of TMP cable
Contact a service center.
End
Breakdown of TMP
Contact a service center.
End
*It is possible to check the details of TMP abnormalities through RS-485 communication.
▶
Troubleshooting