Section 2
Considerations before Installation
2.6.4 Signal
names
Table 2-6
Remote interface signals
Type
S
M
Signal description
Signal name
Input
Y
Y
Compressor start (Momentary operating mode) COMP START
COM
Y Y
Compressor
stop
(
Momentary operating mode
)
COMP STOP
COM
Y Y
Compressor start/stop
(
Alternate operating
mode
)
COMP ON/OFF
COM
Y
Refrigerator No.1start/stop
CH1 REF
COM
Y
Refrigerator No.2start/stop
CH2 REF
Y
Refrigerator No.3start/stop
CH3 REF
Output
Y
Y
Compressor answer-back
COMP RUN1
COMP RUN2
Y Y
Alarm
ALARM1
ALARM2
Y
Refrigerator No.1 answer-back
CH1 ACK1
CH1 ACK2
Y
Refrigerator No.2 answer-back
CH2 ACK1
CH2 ACK2
Y
Refrigerator No.3 answer-back
CH3 ACK1
CH3 ACK2
Note: “S” stands for C30VR as single cryopump operation and “M” stands for C30MVR as multiple
cryopump operation. “Y” indicates available signals for the models. Refer to “Section 4.6 Electrical
connections” for details of signal names.
2.6.5 Signal
functions
The compressor can be controlled in two different operating modes, Momentary operating
mode and Alternate operating mode. You can neither use both operating modes at the
same time, nor switch one operating mode to the other depending on the situations.
Ensure to choose one of the operating modes and connect signal lines needed for the
operating mode you chose. Do not connect signal lines for the operating mode that you
will not use. Decide the operating mode depending on the application of the equipment
side. Momentary operating mode is equivalent to operate by using pushing button switch,
and Alternate operating mode is equivalent to operate by using selector switch.
Alternate operating mode is easier and simpler for use in general.
Selecting operating mode
1. Momentary operating mode
2. Alternate operating mode
Default operating mode is either Momentary operating mode or Alternate operating mode.
The operating mode is switched automatically.The operating mode is determined by
recognizing into which input pin the operation command of the compressor has entered
after turning on the power supply.
Once the operating mode is selected, the mode will be kept until the compressor is turned
off.
ULVAC CRYOGENICS INCORPORATED
2-9
Содержание C30VR Series
Страница 2: ......
Страница 8: ...F 6 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Страница 16: ...S 4 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Страница 18: ...IW 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Страница 32: ...C30VR Compressor Instruction Manual 1 14 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Страница 108: ...CS 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Страница 110: ...Revision History R 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...