Section 6 Wiring and Piping Removal and Storage
place chosen must be with good ventilation to avoid suffocation in case of unexpected
helium leak.
Place the compressor on a flat floor and lock all the casters.
Remove the piping and wiring from the compressor and put dust caps or dust plugs on
the helium line connectors.
Protect the removed components such as flexible hoses and cables with covers or boxes.
Store them with the compressor.
Ensure that cooling water has been completely drained out from the compressor.
To store the compressor, cover it with a plastic sheet or package cardboard. Before
storing the compressor, write down the value of the helium pressure gauge. Check the
value and record it once a month. If the value is clearly decreasing compared to the one
before, there is a possibility of helium leakage. Please contact our Service Engineering
Division or the nearest customer support center.