Page 9
37370 Instruction Manual (Rev. 0)
4.5.3
Mains Cord
It is a standard cable, Cat. # E-WP008, see paragraph 12-ORDERING INFORMATION.
Make sure your power outtake is provided with a reliable ground connection.
4.6 Connections
Connect the mains cord to a power outtake, provided with a reliable earth connection,
protected by a differential earth-leakage switch (CGFI) that breaks at the threshold of
0.03A, within a max. acceptable delay of 1s. The maximum breaking compliance required
by the CGFI is 10kA at rated current is 6A.
4.6.1 Back
Panel
Beside the mains module, the back panel embodies a connection module, see figure be-
low.
Figure 4 “Back Panel”
The connection module encompasses, from left to right:
USB
this USB type B connector for linking the Plantar Test to the USB port of the
PC
COM1
a delta 9-pin female connector for connection to the serial port RS232C of the
PC
COM2
a second delta 9-pin connector (male) for connection to an external serial
printer
MINI
PRINTER
a 20-pin connector to the Mini Printer Cat. 57145, see also (see paragraphs
3.3 and 12.3
TO I.R.
EMITTER
a 7-pin circular connector to link the I.R. Emitter. This cable connector features
a push-pull latch. Pull out the knurled ring to remove the connector.
Mismatching the connection is prevented by the different individual pin arrangement.