Page 23
37370 Instruction Manual (Rev. 0)
Depress the
F1
(
UP
) or
F2
(
DWN
) key to reach the value indicated on the Radiometer,
21mW/cm
2
, then depress
F3
(
STP
) to save the calibration and confirm it by the
F4
(
OK
)
key.
Now depress
F3
(
MED
) key, then
F3
(
STR
).
Depress the
F1
(
UP
) or
F2
(
DWN
) key to
reach the value indicated on the Radiome-
ter,
190mW/cm
2
, then save and confirm it
by the
F3
(
STP
) and
F4
(
OK
) keys.
9.3
Calibration without I.R. Radiometer
Calibrating the Plantar Test without the I.R. Radiometer, means to recall the default cali-
bration value set at our factory during the testing phase.
From the Calibration Menu, depress the
F1
(
DEF
) key and proceed by the
F3
(
YES
) key,
to confirm the change.
The default calibration value is thus recalled.
10 MAINTENANCE
While any service of the instrument ought to be carried out by Ugo Basile personnel or by
qualified personnel authorized by UGO BASILE organization, this section of the instruc-
tion manuals describes normal maintenance procedures which can be carried out at your
facility.