Page 18
37370 Instruction Manual (Rev. 0)
Beside the column
DATE
, which indicates date and time, you find the following columns:
TYPE
indicating the type of instrument,
PLANTEST
, in our case
IDENT
indicating the serial number of the instrument in use, in our case
PT99999
,
where
PT
stays for Plantar Test
and the column
ACQUIRED DATA
, which lists the experimental datum proper, in the fol-
lowing format:
05.0;45;1;M;01;Y;
where:
05.0
= Reaction time
45
= I.R. Intensity
1
= Experiment number
M
= Animal gender
01
= Sequence number
Y
= Test OK
6.4
Quick setting of the 37370 Plantar Test
This Instruction Manual extensively explains how to take advantage of all the features of-
fered by this microprocessor controlled unit. The following table summarizes them; how-
ever, we recommend you to browse the complete manual to go deep into details.
Each sequence of commands is described starting from the
MAIN MENU
.
Command Description
Command Sequence
Paragraph
reference
Set
Printer ON
(FNC)
-
(OUT)
-
(PRN)
6.2
Print DATA
previously saved
in the 37370 memory
(MEM)
-
(OUT)
-
(MPR)
7
Send DATA
previously saved
in the 37370 memory to
memory key
(MEM)
-
(OUT)
-
(MSK)
7
Send DATA
previously saved
in the 37370 memory to
PC
from
CUB
program:
Data/Mem Stick – send all data in mem
YES
Set
DATE & TIME
(FNC)
-
(CLK)
-
(UP)/(DWN) - ( > )
-
(ESC)
7.2
Set
DISPLAY Contrast
(FNC)
-
(DIS)
-
(UP)/(DWN)
depressed
until the desired contrast is reached -
(ESC)
-
(ESC)
5.2
Содержание 37370
Страница 39: ......