Page 14
37370 Instruction Manual (Rev. 0)
5.6.2 Experiment
Number
As soon the
F4
(
OK
) or
F3
(
NUL
) keys are
depressed, followed by the
F4
key again,
the second option is highlighted.
The experiment number, from 1 to 9, can
now be selected via the
F1
(
UP
) or
F2
(
DWN
) key and confirmed via the
F4
(
OK
)
key.
IMPORTANT:
The experiment number is also used to label the data export file on the USB
pen-drive (for example:
DATA01.txt
, where the experiment number is 1,
DA-
TA02.txt
, where the experiment number is 2, etc.
5.6.3 Animal
Gender
After the experiment number, it is possible
to define the gender of the animal to be
processed:
MALE/FEMALE
.
Select it by depressing the
F1
(
UP
) or
F2
(
DWN
) key and confirm it via the
F4
(
OK
)
key; avoid the gender selection by depress-
ing the
F3
.
5.6.4 Cut-Off
Time
To get over a “false-start” and to avoid useless overheating of the paw when the animal
does not react within the expected reaction time, a cut-off time can be set. The preset
value indicates the time (in seconds) after which the experiment will be interrupted.
Select the cut-off time by depressing the
F1
(
OPR
),
F2
(
SET
) and
F2
(
CUT
).
Depress the
F1
(
UP
) or
F2
(
DWN
) key un-
til you reach the desired value and confirm
it via the
F4
(
OK
) key.
5.7 Configuration
File
Up to 6 different experiments can be configured, following the instructions outlined in
previous paragraphs, saved in the memory key, ready to be recalled to the PC and/or to
the instrument.
Содержание 37370
Страница 39: ......