Page 11
37370 Instruction Manual (Rev. 0)
incorrect or improper use or, in any case, not in accordance with the purpose
for which the instrument has been designed and the warnings stated in the in-
struction manual supplied with the instrument;
replacement of original components, accessories or parts with others not ap-
proved by the manufacturer;
servicing carried out by unauthorized personnel
5 OPERATION
5.1 Switching
On
Switch on the 37370; the software takes some seconds to perform a complete checking
of the electronic circuit. During the checking, the display shows the mice logo. By de-
pressing ESC, the display shows the software version which is actually installed, e.g.,
“V1.0”, see paragraph 2.3.2.
The software version is necessary in case of communication with our laboratory for oper-
ational or servicing problems.
In case one of the captions
MEMORY LOSS!!
appears on the display, please refer to para-
graph 10.1.1-Battery Replacement.
5.2 Display
Contrast
To modify the display contrast, depress the
F2 (FNC)
key from the Main Menu, then the
F3 (DIS)
key. The display shows:
Keep the arrow keys
F1 (UP )
or
F2 (DWN)
depressed until you reach the desired dis-
play contrast.
The readability of the liquid crystal display
depends on the angle of view: select the
ideal contrast according to the height of the
table, the operator’s distance and so on.
5.3
Setting I.R. Intensity
Before carrying out the test on the animal,
set the I.R. intensity. Enter the I.R. setting
menu, by depressing
F1
(
OPR
),
F2
(
SET
)
and
F1
(
IR
).
The following display will appear:
Now use the
F1
(
UP
) or
F2
(
DWN
) keys to set intensity to the desired value.