7
LRM (LONG RANGE MODULE).
A cone shaped
scanning module that increases the detection
distance above standard scanning modules. The
LRM-15 is ideal for high voltage inspection and for
locating leaks at great distances.
REMOTE MAGNETIC TRANSDUCER.
The RMT is
a magnetically mountable contact probe with cable.
The probe is applied to a test surface and the RAM
(Remote Access Module) is plugged into the front end
of the Ultraprobe.
PISTOL-GRIP HOUSING
DISPLAY PANEL.
In the Operation Mode, the Display Panel will show inspection parameters such
as intensity levels (as dB and as a bar graph), Frequency, Battery Level, Sensitivity Level, Sensitivity
Value, “Display Mode” and Function Select Field. Intensity levels are sho
wn simultaneously as a
numeric dB value and on a sixteen-segmented bar graph (with each segment representing 3 dB). The
Frequency is adjustable from 20 kHz to 100 kHz. These represent the range of frequency selection
capable with the Ultraprobe. The most common frequency used for general leak detection or electrical
inspection is 40 kHz. These frequencies may be "tuned in" when performing inspections with any of
the Ultraprobe’s plug
-in modules. The Display Mode indicates the operation mode of the instrument.
This is indicated on the display panel as
RT
for
Real Time
,
PH
for
Peak Hold
,
SS
for
Snap Shot
,
or
RTO
for
Real Time “Offset”
,
PHO
for
Peak Hold “Offset”
and
SSO
for
Snap Shot
“Offset”
. (For additional information about Offset see
dB Scale Select
in the Set-Up Mode
Section.
TRIGGER ON/OFF TRIGGER SWITCH.
The Ultraprobe is always "off' until the trigger switch is
pulled. To operate, pull and hold the trigger. To turn the instrument off, release the trigger. The time it
Yellow Enter Button
Sensitivity Control Dial
Display Panel