38
TO INPUT THE DATA
NOTE:
Do not connect any external source into Line Input jack at this point.
Enter Setup Mode
Spin to Menu 20: “Line Input”
Click the Sensitivity dial in to enter
Disabled will blink, Spin to Temp Probe
Click Sensitivity Dial to set; Press Enter to exit.
WHEN READY TO ENTER TEMPERATURE
Connect the cable from the Thermometer to the Ultraprobe Line Input jack.
Click the Sensitivity Dial until the cursor moves do the bottom of the Display Panel
until Press ENTER for is blinking.
Spin until the specific Application info is blinking (e.g.: Generic or Steam).
Press Enter to enter the Application info screen.
Spin to Temperature.
Turn on the Thermometer and take the temperature measurement and hold the
reading.
Click (the temperature numerals will blink rapidly) and the data will be entered.
(Optional) Click to re-enter the data.
Press Enter to save: the temperature will be stored.
NOTE:
If you wish to move the data to another location the store screen will prompt
you to spin to the desired location and then to press Enter again.