Spin the Sensitivity Control Dial until the desired setting (Real Time, Snap Shot or
Peak Hold) appears and blinks.
Click the Sensitivity Dial to set and return to Setup Mode.
Press the Yellow ENTER button to exit.
MENU 29: dB SCALE TYPE SELECT (RELATIVE OR OFFSET)
dB Select has two settings from which to choose. These settings will determine the
baseline dB reference level of the instrument. Once selected, all test results will be based
on the selected baseline dB level.
There are two scales: Relative and dB offset.
RELATIVE
sets the instrument to the 0 dB of the instrument’s internal minimal detection
value and is the factory default setting.
dB OFFSET
is a dB level that is a new minimum reference level set by the user. This
value may be any dB level above the natural 0 dB of the instrument. Once set, the preset
level must be subtracted from the reading to determine an accurate dB increase. (EG: if
“10” is the dB offset value and a subsequent reading is 25 dB, then the increase is 15
dB.)
TO SELECT A dB REFERENCE SCALE
Enter the Setup Mode
Spin to dB Scale Select (Menu 31 blinks).
Click in the Sensitivity Control Dial.
Spin the Sensitivity Control Dial to the desired scale (Relative or Offset).
Click the Sensitivity Control Dial to set.
Press the Yellow ENTER button to exit.