background image

PAGE 10 OF 17 

0921 IH-9560

MENU SETTINGS CONTINUED 

CALIBRATION MENU

ENTER THIS MENU TO PERFORM CALIBRATIONS.

Internal Calibration

Perform

Calibration Adjustment

 -100…0…100

Span:

Perform

Linearity:

Perform

End Calibration:

Exit menu

SPAN [

SPAN

Initiates a span calibration procedure (zero and span). 

LIN [

LIN

Initiates a linearity calibration procedure (zero, 

mid-point and span). 

END CAL [

End

Advance to the next menu or return to the top of the 

current menu.

SETUP MENU

ENTER THIS MENU TO SET SCALE PARAMETERS.

Reset

no, yes

Auxiliary Graduation

On, Off

Filter

Low, Med, High

Auto Zero Tracking

off, 0.5d, 1d, 3d

Stable

0.5d, 1d, 2d, 5d

Backlight

off, on, auto

Auto Tare

off, on, on-acc

Auto Off

off, 1, 5, 10

End Setup

Exit menu

NOTE: Bold always represents factory default. 

RESET [

rESEt

Reset the setup menu to factory defaults. 

•  NO = Not reset 

•  YES = Reset 

AUXILIARY GRADUATION [

Au.GrAd

Set whether the weight is displayed with the auxiliary 

graduation when the security switch is in the locked 

position. 

•  ON = The auxiliary graduation is displayed within 

brackets. 

•  OFF = The auxiliary graduation is not displayed. 

NOTE: If the scale will be used in direct sales to 
the public, the OFF setting must be selected. 

FILTER 

Set the amount of signal filtering. 

•  LOW = Less stability, faster stabilization time

•  MED = Normal stability, stabilization time

•  HI = Greater stability, slower stabilization time

AUTO ZERO TRACKING 

Set the automatic zero tracking functionality. 

•  Off = Disabled 

•  0.5d = The display will maintain zero until a change 

of 0.5 divisions per second has been exceeded. 

•  1d = The display will maintain zero until a change of 

one division per second has been exceeded. 

•  3d = The display will maintain zero until a change of 

three divisions per second has been exceeded. 

STABLE RANGE 

Set the amount the reading can vary while the stability 

symbol remains on.

•  0.5d = 0.5 scale division 

•  1d = 1 scale division

•  2d = 2 scale division

•  5d = 5 scale division

BACKLIGHT 

Sets backlight functionality. 

•  Off = Always off 

•  On = Always on 

•  Auto = Turns on when a button is pressed or the 

displayed weight changes. 

NOTE: When powered by AC adapter, the 
backlight is always on. 

AUTO TARE 

Sets the automatic tare functionality. 

•  Off = Automatic tare is disabled. 

•  On = The first stable gross weight is tared. 

•  On-acc = Stable gross loads within the accept limits 

are tared (in Check Weighing mode). 

Содержание SCOUT H-9560

Страница 1: ...WARNING SYMBOLS General Hazard Electric Shock Hazard CAUTION Read all safety warnings before installing making connections or servicing this equipment Failure to comply with these warnings could resul...

Страница 2: ...cale RELEASING TRANSPORTATION LOCK Release the red transportation lock by turning the red pointer 90 degrees counterclockwise to unlock See Figure 1 INSTALLING THE WEIGHING PAN For scales with a recta...

Страница 3: ...and connect the AC plug to an electrical outlet See Figure 5 BATTERY INSTALLATION Install the four AA batteries with polarity as shown in the battery compartment See Figure 6 NOTE After powering on th...

Страница 4: ...f Function Initiates an application mode Tare Enter clear a tare value Secondary Function Long Press Extended Press Zero Off Turns off the scale None Units Changes the weighing unit None Mode Allows c...

Страница 5: ...5 6 7 8 9 10 11 12 13 ITEM DESCRIPTION 1 Battery charge 2 Stable weight 3 Negative 4 Center of Zero 5 NET 6 Preset Tare 7 Pointers 8 Kilogram grain 9 Check Weighing 10 Pound Ounce Pound Ounce Symbols...

Страница 6: ...and hold Mode until Count is displayed CLr PW clear average piece weight APW will then display If no APW exists the scale will display PUt 10 proceed to step 5 3 Press No to use the stored APW Proceed...

Страница 7: ...an the Over limit Place the item to be weighed on the scale and press Tare Remove a portion of the item until it is within the Accept range ZERO CHECK Zero check is used when comparing subsequent samp...

Страница 8: ...two modes for the display hold Peak Hold Allows the user to capture and store the highest stable weight value 5d Display Hold default Allows the user to capture and store the first stable weight valu...

Страница 9: ...On Off Totalize On Off Density On Off Hold Disp Peak Off Reset g kg ct oz ozt dwt lb grn hkt sgt twt t tola tical End End End End End End Available units may vary by local regulations Available settin...

Страница 10: ...y graduation is not displayed NOTE If the scale will be used in direct sales to the public the OFF setting must be selected FILTER Set the amount of signal filtering LOW Less stability faster stabiliz...

Страница 11: ...et the Mode menu to factory defaults No Not reset Yes Reset WEIGH WEIGH Set the status Off Disabled On Enabled COUNT Count Set the status Off Disabled On Enabled PERCENT PErcnt Set the status Off Disa...

Страница 12: ...ore turning the scale over remove the pan and pan support if present and set the transportation lock to locked position The scale can be supported using lab jacks or any other convenient method Ensure...

Страница 13: ...8 5 Tare out of range Err 8 6 Displayed value exceeds 999999 possible in Totalization mode Err 8 7 Unknown internal weight position rEF Err Parts counting or percentage error sample weight 1d Scale sh...

Страница 14: ...ocal weights and measures office for more information SEALING After the unit has been verified it must be sealed by the weights and measures official to prevent undetected access to the legally contro...

Страница 15: ...ernal span or linearity calibration Tare Range Full Capacity by subtraction Weighing Units g kg ct oz ozt dwt lb grn Application Modes Weighing Parts Counting Percent Weighing Check Weighing Totalizat...

Страница 16: ...andards of EU Directives 2011 65 EU RoHS 2014 30 EU EMC 2014 35 EU LVD and 2014 31 EU NAWI The complete Declaration of Conformity is available online at www ohaus com ce This product complies with the...

Страница 17: ...erference in a residential installation This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interfe...

Отзывы: