PAGE 14 OF 17 

0921 IH-9560

LEGAL FOR TRADE

TECHNICAL DATA

When the scale is used in trade or a legally controlled 

application it must be set up, verified and sealed 

in accordance with local weights and measures 

regulations. It is the responsibility of the purchaser to 

ensure that all pertinent legal requirements are met. As 

the requirements vary by jurisdiction, the purchaser is 

advised to contact local weights and measures office 

for instructions about putting the scale into service.

SETTINGS 

Before verification and sealing, perform the following 

steps: 

1.  Verify that the menu settings meet the local weights 

and measures regulations. 

2.  Units menu should be reviewed. Verify the units 

turned on are permitted by the local weights and 

measures regulations. 

3.  Perform a calibration. 

4.  Set the security switch to the locked position.

VERIFICATION

A weights and measures official must perform the 

verification procedure. Contact the local weights and 

measures office for more information.

SEALING

After the unit has been verified, it must be sealed by the 

weights and measures official to prevent undetected 

access to the legally controlled settings. Before sealing 

the device, ensure that the security switch is in the 

locked position. (See Figure 9)

If using a wire seal, pass the sealing wire through the 

holes in the security switch and bottom housing.  

(See Figure 9)

If using a paper seal, place the seal over the security 

switch and bottom housing. (See Figure 9)

Figure 9

Unlocked

Locked With Paper Seal

Locked With Wire Seal

Equipment Ratings: Indoor use only

Altitude: Up to 6,560 ft

Operating temperature range: 41°F to 104°F

Specified temperature range: Refer to tables on  

page 15

Relative humidity: 10% to 80% at 88°F, decreasing 

linearly to 50% at 104°F, non-condensing

Power: AC power adapter input 100-240V 50/60 Hz and 

output 5 V DC 1 A (for use with certified or approved 

power supply, which must have a SELV and limited 

energy circuit output), or 4 AA batteries

Main supply voltage fluctuations: Up to +/- 10% of the 

nominal voltage

Pollution degree: 2

Installation category: II

Содержание SCOUT H-9560

Страница 1: ...WARNING SYMBOLS General Hazard Electric Shock Hazard CAUTION Read all safety warnings before installing making connections or servicing this equipment Failure to comply with these warnings could resul...

Страница 2: ...cale RELEASING TRANSPORTATION LOCK Release the red transportation lock by turning the red pointer 90 degrees counterclockwise to unlock See Figure 1 INSTALLING THE WEIGHING PAN For scales with a recta...

Страница 3: ...and connect the AC plug to an electrical outlet See Figure 5 BATTERY INSTALLATION Install the four AA batteries with polarity as shown in the battery compartment See Figure 6 NOTE After powering on th...

Страница 4: ...f Function Initiates an application mode Tare Enter clear a tare value Secondary Function Long Press Extended Press Zero Off Turns off the scale None Units Changes the weighing unit None Mode Allows c...

Страница 5: ...5 6 7 8 9 10 11 12 13 ITEM DESCRIPTION 1 Battery charge 2 Stable weight 3 Negative 4 Center of Zero 5 NET 6 Preset Tare 7 Pointers 8 Kilogram grain 9 Check Weighing 10 Pound Ounce Pound Ounce Symbols...

Страница 6: ...and hold Mode until Count is displayed CLr PW clear average piece weight APW will then display If no APW exists the scale will display PUt 10 proceed to step 5 3 Press No to use the stored APW Proceed...

Страница 7: ...an the Over limit Place the item to be weighed on the scale and press Tare Remove a portion of the item until it is within the Accept range ZERO CHECK Zero check is used when comparing subsequent samp...

Страница 8: ...two modes for the display hold Peak Hold Allows the user to capture and store the highest stable weight value 5d Display Hold default Allows the user to capture and store the first stable weight valu...

Страница 9: ...On Off Totalize On Off Density On Off Hold Disp Peak Off Reset g kg ct oz ozt dwt lb grn hkt sgt twt t tola tical End End End End End End Available units may vary by local regulations Available settin...

Страница 10: ...y graduation is not displayed NOTE If the scale will be used in direct sales to the public the OFF setting must be selected FILTER Set the amount of signal filtering LOW Less stability faster stabiliz...

Страница 11: ...et the Mode menu to factory defaults No Not reset Yes Reset WEIGH WEIGH Set the status Off Disabled On Enabled COUNT Count Set the status Off Disabled On Enabled PERCENT PErcnt Set the status Off Disa...

Страница 12: ...ore turning the scale over remove the pan and pan support if present and set the transportation lock to locked position The scale can be supported using lab jacks or any other convenient method Ensure...

Страница 13: ...8 5 Tare out of range Err 8 6 Displayed value exceeds 999999 possible in Totalization mode Err 8 7 Unknown internal weight position rEF Err Parts counting or percentage error sample weight 1d Scale sh...

Страница 14: ...ocal weights and measures office for more information SEALING After the unit has been verified it must be sealed by the weights and measures official to prevent undetected access to the legally contro...

Страница 15: ...ernal span or linearity calibration Tare Range Full Capacity by subtraction Weighing Units g kg ct oz ozt dwt lb grn Application Modes Weighing Parts Counting Percent Weighing Check Weighing Totalizat...

Страница 16: ...andards of EU Directives 2011 65 EU RoHS 2014 30 EU EMC 2014 35 EU LVD and 2014 31 EU NAWI The complete Declaration of Conformity is available online at www ohaus com ce This product complies with the...

Страница 17: ...erference in a residential installation This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interfe...

Отзывы: