PAGE 8 OF 17
0921 IH-9560
OPERATION CONTINUED
PREPARATION
Remove the pan, insert the hook and install the beaker
stand (beaker is not included). Press the Zero button to
turn on the scale. (See Figure 6)
DENSITY SETUP
1. Press and hold Mode until [
SPEC.Gr
] is displayed.
[
Air.Wt
] will then be displayed.
2. Place the sample on the hook as shown in Figure 6
and press Yes to store the dry weight value. [
W.Wt
]
will then be displayed.
3. Suspend the sample in water and press Yes to store
the wet weight value. The specific gravity value will
then be displayed and the Gravity icon will blink
while the unit icon is turned off.
NOTE: Specific gravity has no unit of measure.
For large samples, it is recommended to use
the "weighing below" feature instead of the
hook. (See Weight Below Hook on page 12)
HOLD MODE
There are two modes for the display hold:
• Peak Hold: Allows the user to capture and store the
highest stable weight value (>=5d).
• Display Hold (default): Allows the user to capture
and store the first stable weight value (>=5d).
START
If no weight value is held on the display, press Function
key to begin. [
rEAdY
] will be displayed until a weight is
added on the pan.
When the stable value is being held on the display, the
Hold icon will blink and the displayed weight will not
change.
RESET
If the pan is empty and a weight value was held on the
display, a single short press of the Function key will clear
the held value and show the new weight on the pan.
1. Press and hold Mode until [
HoLd
] is displayed.
2. If required, place an empty container on the pan
and press Tare. Zero value will then display.
3. Press Function key to begin. [
rEAdY
] will be displayed.
4. Place samples to be weighed on the pan.
5. The stable value will be held on the display. The
Hold icon will blink.
Figure 6
Beaker Stand
Preparation
Air Weight
Wet Weight