IEC 61000-4-2
floors are covered with
synthetic material, relative
humidity should be at least
30%.
Electrical Fast
Transient/Burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for
input/output lines
± 2 kV for power supply
lines
Not applicable
Mains power quality should be
that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1 kV differential
mode
± 2 kV common
mode
± 1 kV differential
mode
Not applicable
Mains power quality should be
that of a typical
commercial or hospital
environment.
Voltage dips, short
interruptions,
and voltage variations on
power
supply input lines
IEC 61000-4-11
< 5 % U
T
(> 95 % dip in U
T
)
für 0.5 cycles
70 % U
T
(30 % dip in U
T
) for
25 cycles
< 5 % U
T
(> 95 % dip in U
T
)
for 5 s
< 5 % U
T
(> 95 dip in U
T
) for 0.5
cycles
70 % U
T
(30 % dip in U
T
) for 25
cycles
< 5 % U
T
(> 95 % dip in U
T
) for 5
s
Mains power quality should be
that of a typical
commercial or hospital
environment.
Power Frequency
(50/60 Hz) Magnetic Field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
NOTE: U
T
is the AC mains voltage before application of the test level.
Table 3: Guidelines and Manufacturers‘s declaration – Electromagnetic Immunity
Guidance and Manufacturer’s Declaration— ELECTROMAGNETIC
IMMUNITY
The PABLO® system is designed for operation in an ELECTROMAGNETIC ENVIRONMENT as indicated
below. The customer or user of the PABLO® system must ensure that it is used in such an environment.
Portable and mobile RF communications equipment should be used no closer to any part of the device,
including cables, than the recommended separation distance calculated from the equation applicable to
the frequency of the transmitter.