EN – 17
Measurement deviation:
Deviation force measurement: < ±5% ± 2 LSB (1 - 20N)
Deviation force measurement: < ±2% ± 2 LSB (20 - 1000N)
Operational area: 10 meters given uninterrupted view of computer device
Penetration protection:
IP22
Weight:
PABLO® Handsensor (in g): 345
PABLO® Motionsensor (in g): 40
PABLO® Ladestation (in g): 950
PABLO® Multiball (in g): 750
PABLO® Multiboard (in g): 1600
Dimensions (LxWxH):
PABLO® Handsensor (in mm): 200 x 68 x 39
PABLO® Motionsensor (in mm): 56 x 34 x 21
PABLO® Charger (in mm): 210 x 180 x 37
PABLO® Multiball (in mm): Ø 140
PABLO® Multiboard (in mm): 520 x 338 x 257
System Requirements:
•
Windows 7, Windows 8 or Windows 10; Home or Professional
•
Intel or AMD Processor with >= 1.6GHz
•
at least 4 GB RAM
•
Display resolution 1280x768
•
Microsoft .NET framework 4.5.2 (installed by default)
•
Unity3D Webplayer Plugin (installed by default)
One free USB port
2.1.2
Area of application
Device usage is principally limited to clean, dry rooms within professional health care
establishments.
Operation:
Temperature: 10 ... 30 °C
Humidity: 30 ... 75 % relative humidity
Storage and transport:
Temperature: -20 ... 60 °C
Humidity: 20 ... 90 % relative humidity, no dew
Caution:
The PABLO® system may not be used in explosion-prone zones AP and APG according to
EN 60601-1/2006.
Among other things, this means:
The usage of easily flammable and explosive, anesthetic inhalation materials and
mixtures thereof is not permitted within the vicinity of the PABLO® system. These
materials include:
• Ether pro narcosi (diethyl ether)
• Cyclopropane