EN – 20
Immunity Test
IEC 60601-Test
Level
Compliance Level
ELECTROMAGNETIC
ENVIRONMENT - Guidance
Recommended Separation
Distance:
Conducted RF
IEC 61000-4-6
6 V
rms
150 kHz to 80 MHz
6 V
𝑑 = 0.58 √𝑃
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to 2.7 GHz
10 V/m
𝑑 = 0.35 √𝑃
80 MHz to 800
MHz
𝑑 = 0.7 √𝑃
800 MHz to 2.5
GHz
where P is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site
survey
a
, should be less than the
compliance level in each
frequency range.
b
Interference may occur in the
vicinity of equipment marked
with the following symbol:
NOTES:
- At 80 MHz and 800 MHz, the higher frequency range applies.
- These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects, and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in
the location in which the PABLO® system is used exceeds the applicable RF compliance level above,